Notifications
You must be signed in to change notification settings - Fork 38
Expand file tree
/
Copy pathpfeiffer.py
More file actions
245 lines (207 loc) · 8.12 KB
/
pfeiffer.py
File metadata and controls
245 lines (207 loc) · 8.12 KB
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
176
177
178
179
180
181
182
183
184
185
186
187
188
189
190
191
192
193
194
195
196
197
198
199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
218
219
220
221
222
223
224
225
226
227
228
229
230
231
232
233
234
235
236
237
238
239
240
241
242
243
244
245
"""This module contains drivers for the following equipment from Pfeiffer
Vacuum:
* TPG 262 and TPG 261 Dual Gauge. Dual-Channel Measurement and Control
Unit for Compact Gauges
"""
import time
import serial
# Code translations constants
MEASUREMENT_STATUS = {
0: 'Measurement data okay',
1: 'Underrange',
2: 'Overrange',
3: 'Sensor error',
4: 'Sensor off (IKR, PKR, IMR, PBR)',
5: 'No sensor (output: 5,2.0000E-2 [mbar])',
6: 'Identification error',
}
GAUGE_IDS = {
'TPR': 'Pirani Gauge or Pirani Capacitive gauge',
'IKR9': 'Cold Cathode Gauge 10E-9 ',
'IKR11': 'Cold Cathode Gauge 10E-11 ',
'PKR': 'FullRange CC Gauge',
'PBR': 'FullRange BA Gauge',
'IMR': 'Pirani / High Pressure Gauge',
'CMR': 'Linear gauge',
'noSEn': 'no SEnsor',
'noid': 'no identifier',
}
PRESSURE_UNITS = {0: 'mbar/bar', 1: 'Torr', 2: 'Pascal'}
class TPG26x(object):
r"""Abstract class that implements the common driver for the TPG 261 and
TPG 262 dual channel measurement and control unit. The driver implements
the following 6 commands out the 39 in the specification:
* PNR: Program number (firmware version)
* PR[1,2]: Pressure measurement (measurement data) gauge [1, 2]
* PRX: Pressure measurement (measurement data) gauge 1 and 2
* TID: Transmitter identification (gauge identification)
* UNI: Pressure unit
* RST: RS232 test
This class also contains the following class variables, for the specific
characters that are used in the communication:
:var ETX: End text (Ctrl-c), chr(3), \\x15
:var CR: Carriage return, chr(13), \\r
:var LF: Line feed, chr(10), \\n
:var ENQ: Enquiry, chr(5), \\x05
:var ACK: Acknowledge, chr(6), \\x06
:var NAK: Negative acknowledge, chr(21), \\x15
"""
ETX = chr(3) # \x03
CR = chr(13)
LF = chr(10)
ENQ = chr(5) # \x05
ACK = chr(6) # \x06
NAK = chr(21) # \x15
def __init__(self, port='/dev/ttyUSB0', baudrate=9600):
"""Initialize internal variables and serial connection
:param port: The COM port to open. See the documentation for
`pyserial <http://pyserial.sourceforge.net/>`_ for an explanation
of the possible value. The default value is '/dev/ttyUSB0'.
:type port: str or int
:param baudrate: 9600, 19200, 38400 where 9600 is the default
:type baudrate: int
"""
# The serial connection should be setup with the following parameters:
# 1 start bit, 8 data bits, No parity bit, 1 stop bit, no hardware
# handshake. These are all default for Serial and therefore not input
# below
self.serial = serial.Serial(port=port, baudrate=baudrate, timeout=1)
def _cr_lf(self, string):
"""Pad carriage return and line feed to a string
:param string: String to pad
:type string: str
:returns: the padded string
:rtype: str
"""
return string + self.CR + self.LF
def _send_command(self, command):
"""Send a command and check if it is positively acknowledged
:param command: The command to send
:type command: str
:raises IOError: if the negative acknowledged or a unknown response
is returned
"""
self.serial.write(self._cr_lf(command))
response = self.serial.readline()
if response == self._cr_lf(self.NAK):
message = 'Serial communication returned negative acknowledge'
raise IOError(message)
elif response != self._cr_lf(self.ACK):
message = 'Serial communication returned unknown response:\n{}' ''.format(
repr(response)
)
raise IOError(message)
def _get_data(self):
"""Get the data that is ready on the device
:returns: the raw data
:rtype:str
"""
self.serial.write(self.ENQ)
data = self.serial.readline()
return data.rstrip(self.LF).rstrip(self.CR)
def _clear_output_buffer(self):
"""Clear the output buffer"""
time.sleep(0.1)
just_read = 'start value'
out = ''
while just_read != '':
just_read = self.serial.read()
out += just_read
return out
def program_number(self):
"""Return the firmware version
:returns: the firmware version
:rtype: str
"""
self._send_command('PNR')
return self._get_data()
def pressure_gauge(self, gauge=1):
"""Return the pressure measured by gauge X
:param gauge: The gauge number, 1 or 2
:type gauge: int
:raises ValueError: if gauge is not 1 or 2
:return: (value, (status_code, status_message))
:rtype: tuple
"""
if gauge not in [1, 2]:
message = 'The input gauge number can only be 1 or 2'
raise ValueError(message)
self._send_command('PR' + str(gauge))
reply = self._get_data()
status_code = int(reply.split(',')[0])
value = float(reply.split(',')[1])
return value, (status_code, MEASUREMENT_STATUS[status_code])
def pressure_gauges(self):
"""Return the pressures measured by the gauges
:return: (value1, (status_code1, status_message1), value2,
(status_code2, status_message2))
:rtype: tuple
"""
self._send_command('PRX')
reply = self._get_data()
# The reply is on the form: x,sx.xxxxEsxx,y,sy.yyyyEsyy
status_code1 = int(reply.split(',')[0])
value1 = float(reply.split(',')[1])
status_code2 = int(reply.split(',')[2])
value2 = float(reply.split(',')[3])
return (
value1,
(status_code1, MEASUREMENT_STATUS[status_code1]),
value2,
(status_code2, MEASUREMENT_STATUS[status_code2]),
)
def gauge_identification(self):
"""Return the gauge identication
:return: (id_code_1, id_1, id_code_2, id_2)
:rtype: tuple
"""
self._send_command('TID')
reply = self._get_data()
id1, id2 = reply.split(',')
return id1, GAUGE_IDS[id1], id2, GAUGE_IDS[id2]
def pressure_unit(self):
"""Return the pressure unit
:return: the pressure unit
:rtype: str
"""
self._send_command('UNI')
unit_code = int(self._get_data())
return PRESSURE_UNITS[unit_code]
def rs232_communication_test(self):
"""RS232 communication test
:return: the status of the communication test
:rtype: bool
"""
self._send_command('RST')
self.serial.write(self.ENQ)
self._clear_output_buffer()
test_string_out = ''
for char in 'a1':
self.serial.write(char)
test_string_out += self._get_data().rstrip(self.ENQ)
self._send_command(self.ETX)
return test_string_out == 'a1'
class TPG262(TPG26x):
"""Driver for the TPG 262 dual channel measurement and control unit"""
def __init__(self, port='/dev/ttyUSB0', baudrate=9600):
"""Initialize internal variables and serial connection
:param port: The COM port to open. See the documentation for
`pyserial <http://pyserial.sourceforge.net/>`_ for an explanation
of the possible value. The default value is '/dev/ttyUSB0'.
:type port: str or int
:param baudrate: 9600, 19200, 38400 where 9600 is the default
:type baudrate: int
"""
super(TPG262, self).__init__(port=port, baudrate=baudrate)
class TPG261(TPG26x):
"""Driver for the TPG 261 dual channel measurement and control unit"""
def __init__(self, port='/dev/ttyUSB0', baudrate=9600):
"""Initialize internal variables and serial connection
:param port: The COM port to open. See the documentation for
`pyserial <http://pyserial.sourceforge.net/>`_ for an explanation
of the possible value. The default value is '/dev/ttyUSB0'.
:type port: str or int
:param baudrate: 9600, 19200, 38400 where 9600 is the default
:type baudrate: int
"""
super(TPG261, self).__init__(port=port, baudrate=baudrate)